Measuring nitrate with the colorimeter

Hach NitraVer 5 Reagent

«  Introduction   ::   Contents   ::   API nitrate test kit  »

Hach NitraVer 5 Reagent

_images/nitrate_hach.JPG

Reagents & Equipment

  • Colorimeter with 400 nm LED (part # COLOR-400)
  • Cuvettes (macro or semi-macro)
  • NitraVer 5 Nitrate Reagent for 10 mL samples, Hach cat # 21061-69
  • Stir plate and stir bar for mixing
  • 1 mL pipette
  • Small beaker
  • (Optional) Distilled water for sample dilutions

Colorimeter set-up

  • Wavelength: 400 nm
  • Nitrate standard data: Mid-range works well (0 - 80 ppm NO3). Download a data file or prepare your own nitrate standard data
  • Before measuring blank the colorimeter against water or a distilled water sample developed with NitraVer 5 reagent as described below

Method

  • Transfer 10 mL of water sample to a small beaker with a stir bar and place on stir plate.
  • Add the contents of a packet of NitraVer5 to the beaker and start mixing.
  • After 1 minute, take the beaker off the stir plate. Remove the stir bar.
  • Wait at least 5 mins for the color to develop.
  • Pour some of the sample into a cuvette, place in the colorimeter and click measure. The program will return the nitrate concentration in ppm.